EFT/ Burst Generator PEFT Junior
47
6.3.4.1 Entry format
Given in ASCII characters (7 digit). Block and line terminations use the conventional CR
(Carriage Return), LF (LineFeed), etc.
Source Handshake
SH1
complete capability
Acceptor Handshake
AH1
complete capability
Talker:
T5
Basic type of talker
Series enquiry,
sounds always
faulty
De-addressing
with MLA
Listener:
L3
Basic type of listener Sounds always
faulty
De-addressing
with MTA
Service Request
SR1
complete capability
Remote Local
RL2
no control with
interlock
Parallel Poll
PP0
no capability
Device clear
DC0
no capability
Device Trigger
DT0
no capability
Controller: C0
no
capability
Tab.
6.3.4.2
The bus address MTA ( My Talk Address ) and MLA ( My Listen Address ) are programable in the
range 0 to 30.
6.3.4.2 Entry format
Given in ASCII characters (7 digit). Block and line terminations use the conventional CR
(Carriage Return), LF (LineFeed), etc.
6.3.4.3 Serial Interface RS-232C
This interface uses the TxD and RxD lines for communication. For communication the hardware
RTS/CTS and the software XON/XOFF can be used.
6.3.4.4 Switching remote mode / local mode
Two modes of operation exist (LOCAL) and (REMOTE).
Equipment not fitted with a REMOTE CONTROL type 730, is automatically set to local mode
when primary power is applied.
If the REMOTE CONTROL type 730 is fitted, local mode is set by the command "
GTL
" (
Go To
Local
). Remote operation is selected with the command "
REN
" (
Remote ENable
).
6.3.5 Programing
The P90 controller, can be operated by a remote device. The data is composed of parameter
commands plus handshake information. Control commands sent to the P90 are held in a buffer