AUTION ELEVEN
™
AE-4022
Sample Measurement
5. Confi rm normal measurement icon
AE-4020
2-6
Chapter 2 Measurement Operation
Take out the required number of test strips from
the bottle.
Close the test strip bottle cap securely.
IMPORTANT:
After taking the test strips out
of the bottle, immediately close
the cap or the test strips in the
bottle will absorb moisture or dirt
in the air, which may deteriorate
reagents and make the test
strips useless.
5
Start normal measurement
Confirm that the icon (
) for normal measurement
appears on the [Standby screen].
IMPORTANT:
Different procedures at this
stage are required, depending
on the setting of the “opera-
tional mode when turning ON”
parameter. (See “Operational
mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the “operational
mode when turning ON” parameter, see “3-7-8. No.005: Operational
mode when turning ON” on page 3-8.
REFERENCE:
When the buzzer setting is ON, the buzzer sounds in the following
intervals to signal the dipping timing for the test strips.
Slow beeping
Fast beeping
Preparation for dipping:
3 seconds
Dipping period:
2 seconds
Test strip setting:
.4 seconds
● When using the “Auto start” mode, go to step 6.
● When using the “Cycle start” mode, press and go to step 6.
6
Call up the [Clarity input screen]
Press and hold
for about 2 seconds.
The [Clarity input screen] appears.
appears on standby screen.
6. Enter Clarity Value (If feature is ON).
• Press and hold
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with
buttons
Select
[ Select test strip type ]
[ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
for about 2 seconds.
• Enter the clarity value of sample using numeric keys
• Enter a number between 1 and 5 using table below:
• Press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with
buttons
Select
[ Select test strip type ]
[ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
to confi rm entry.
7. Dip entire test strip pad area in sample for two (2) seconds.
• If buzzer setting is ON follow the timing chart below:
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with
buttons
Select
[ Select test strip type ]
[ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
IMPORTANT:
• Do NOT wet the black
mark on test strip.
• After dipping test strip in
sample, gently remove
excess urine by drawing
the entire length of strip
against the rim of the
collection cup.
AE-4020
2-6
Chapter 2 Measurement Operation
Take out the required number of test strips from
the bottle.
Close the test strip bottle cap securely.
IMPORTANT:
After taking the test strips out
of the bottle, immediately close
the cap or the test strips in the
bottle will absorb moisture or dirt
in the air, which may deteriorate
reagents and make the test
strips useless.
5
Start normal measurement
Confirm that the icon (
) for normal measurement
appears on the [Standby screen].
IMPORTANT:
Different procedures at this
stage are required, depending
on the setting of the “opera-
tional mode when turning ON”
parameter. (See “Operational
mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the “operational
mode when turning ON” parameter, see “3-7-8. No.005: Operational
mode when turning ON” on page 3-8.
REFERENCE:
When the buzzer setting is ON, the buzzer sounds in the following
intervals to signal the dipping timing for the test strips.
Slow beeping
Fast beeping
Preparation for dipping:
3 seconds
Dipping period:
2 seconds
Test strip setting:
.4 seconds
● When using the “Auto start” mode, go to step 6.
● When using the “Cycle start” mode, press and go to step 6.
6
Call up the [Clarity input screen]
Press and hold
for about 2 seconds.
The [Clarity input screen] appears.
1
2
3
4
5
Clear
Slightly
Cloudy
Cloudy
Turbid
Other
2.4 Measurement operation
AE-4022
2-19
6 Call up the [Clarity input screen]
a
Press and hold
for about 2 seconds.
• The [Clarity input screen] appears.
7 Enter the clarity value
a
Enter the clarity value of the sample using the numeric keys.
• Enter a number between 1 and 5 using the table below.
b
Press
.
• The display returns to the [Standby screen].
8 Dip a test strip in the sample for 2 seconds
a
Dip the test strip in the sample for 2 seconds, and then withdraw it.
Input No.
1
2
3
4
5
Clarity
Clear
Slightly
cloudy
Cloudy
Turbid
Other
REFERENCE:
When the clarity value of the sample is “1”, just press
without pressing any numeric keys.
IMPORTANT:
Dip the entire test strip pad area in the sample at once.
Do not
allow the
black mark shown in the figure to become wet. If the test strip is dipped too
deeply and the black mark is moistened, correct measurement may not be
obtained.
After dipping the test strip in the sample, gently remove excess urine by
drawing the entire length of the test strip against the rim of the collection cup.
In the “Auto start” mode, the buzzer does not sound for the first measurement. Always dip the test strip for 2
seconds. Insufficient dipping of test strips may cause insufficient color change, while dipping too long may
cause reagents to drain out of the test strip, either of which will prevent correct measurement.
Do not
dip the
black mark.
Continued on next page…
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...