10
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Maintenance – Daily
9. Wash detached test strip tray, suction port, test strip feed tray and feed lever with water.
10. Sterilize the carrying arm using alcohol. Then wash with water.
11. Wipe the parts with a cloth and dry completely before reattaching to instrument.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Clean the Feeder
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...