![Arkray AUTION ELEVEN AE-4022 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/arkray/aution-eleven-ae-4022/aution-eleven-ae-4022_quick-reference-manual_2972874025.webp)
15
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Errors
Message Error
Possible cause
Measures (Corrective Action)
E007
Data not
found
No data was found in the specifi ed range.
Verify that the specifi ed range for re-printing,
re-sending, or list printing is appropriate.
E008
Auto start
sensor error
• A test strip was placed while the carrying arm
was moving to the suction ports.
• The detection section of the auto start sensor
was soiled with urine.
• The auto start sensor is faulty.
• Remove the test strip.
• Clean the detection section.
• If you fi nd no abnormality and the error recurs,
contact your local distributor.
E009
No test strip
on the feeder
• The feed lever is not attached.
• The incoming strip sensor is faulty.
• The sensor failed to detect a test strip.
• Turn OFF the instrument and open the maintenance
cover.
• Mount the feed lever.
• If you fi nd any obstacles or scattered test strips
inside, remove them.
• If you fi nd no abnormality and the error recurs,
contact your local distributor.
E011
Elapse of QC
deadline
The QC deadline elapsed while the instrument
was OFF or on standby.
Perform control measurements.
Situation
Error
Possible cause
Measures
Buttons will not respond and/or testing does
not start when strip is placed.
Elapse of operator ID
effective time.
The operator ID effective time
elapsed.
Enter operator ID.
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...