Arkray AUTION ELEVEN AE-4022 Скачать руководство пользователя страница 24

14

SEMI-AUTOMATED URINE ANALYZER

AUTION ELEVEN 

AE-4022

Errors

Message Error

Possible cause

Measures (Corrective Action)

E001

Power down

The instrument was suddenly turned OFF 
during measurement.

Retest the sample(s) currently being measured.

E002

Backup 
memory error

As the instrument was unused for a long 
period, the memory backup battery was 
exhausted and the measurement result stored 
in the memory were lost.

Keep the instrument powered for at least 11 hours 
to charge the backup battery. After charging the 
instrument, set the date and time. 

E004

No paper in 
the printer

•  The thermal recording paper has run out.
•  The thermal recording paper roll was not 

properly installed.

•  Install a new roll of thermal recording paper. 
•  Make sure the thermal recording paper is installed 

correctly.

E005

Waste box 
is full

At the end of the measurement, the test 
strip waste counter exceeded 90.

Cancel the error notice by pressing 

4.2    Replacing the thermal recording paper

AE-4022   

4-13

3 Prepare a new roll of thermal recording paper

a

Cut the first (outermost) layer of a new thermal recording paper 
roll, and trim the end to make it straight. A skewed or odd-shaped 
end may cause the thermal recording paper to jam.

4 Set the new thermal recording paper roll

a

Place the new roll of thermal recording paper in the paper holder, 
orienting it correctly so that the paper feeds from the bottom.
Guide the trimmed end of the thermal recording paper into and 
through the slot that receives the recording paper.
The thermal recording paper is automatically wound forward and 
fed into the unit.

NOTE:

When the thermal recording paper is completely removed 
from the printer, the screen message shown in the right 
figure appears and a continuous beep sounds for 
approximately 1 minute. The alarm sound can be stopped 
by pressing 

 .

Paper holder

Set the new roll of thermal 

recording paper in the 

paper holder in the proper 

orientation so that the end 

feeds from the bottom.

NOTE:

Always press the 

 button after replacing the thermal 

recording paper.If you 

do not

 fully feed the thermal 

recording paper through the unit by pressing the 

 button, 

the printer will fail to print the measurement results.

REFERENCE: 

If the printer fails to properly wind and feed the thermal recording paper, press 

 and verify that the paper feeds 

correctly.

. Empty the 

used test strips from the waste box, and press 

.

E006

Surplus urine 
is full

At the end of the measurement, the total 
number of measurements since turning ON the 
instrument exceeded 190.

Turn OFF the instrument and discard any surplus urine 
from the suction port.

Errors may occur when you operate the instrument incorrectly or make a mistake 
during normal operation. When an error occurs, an error message appears on the 
screen and an alarm sounds for approximately 1 minute.
Press 

4.2    Replacing the thermal recording paper

AE-4022   

4-13

3 Prepare a new roll of thermal recording paper

a

Cut the first (outermost) layer of a new thermal recording paper 
roll, and trim the end to make it straight. A skewed or odd-shaped 
end may cause the thermal recording paper to jam.

4 Set the new thermal recording paper roll

a

Place the new roll of thermal recording paper in the paper holder, 
orienting it correctly so that the paper feeds from the bottom.
Guide the trimmed end of the thermal recording paper into and 
through the slot that receives the recording paper.
The thermal recording paper is automatically wound forward and 
fed into the unit.

NOTE:

When the thermal recording paper is completely removed 
from the printer, the screen message shown in the right 
figure appears and a continuous beep sounds for 
approximately 1 minute. The alarm sound can be stopped 
by pressing 

 .

Paper holder

Set the new roll of thermal 

recording paper in the 

paper holder in the proper 

orientation so that the end 

feeds from the bottom.

NOTE:

Always press the 

 button after replacing the thermal 

recording paper.If you 

do not

 fully feed the thermal 

recording paper through the unit by pressing the 

 button, 

the printer will fail to print the measurement results.

REFERENCE: 

If the printer fails to properly wind and feed the thermal recording paper, press 

 and verify that the paper feeds 

correctly.

to cancel the error notice. The display returns to the [Standby screen].

To prevent the error from recurring, take appropriate measures (corrective action), 
referring to the table below. If unable to clear the error, note error code and call 
ARKRAY Technical Support at 1.855.646.3108.

5.2    Error messages

AE-4022   

5-3

5.2

Error messages

Errors may occur when you operate the instrument incorrectly or make a mistake during normal operation.
When an error occurs, an error message appears on the screen and an alarm (pi-pi-pi-pi) sounds for approximately 1 
minute.

Press 

 to cancel the error notice. The display returns to the [Standby screen].

To prevent the error from recurring, take appropriate measures, referring to the table below. Alternatively, note the 
details of the situation at hand, turn OFF the instrument, and contact your local distributor.

Message

Error

Possible cause

Measures

E001

Power down

The instrument was suddenly turned 

OFF during measurement.

Retest the sample(s) currently being 

measured.

E002

Backup memory error

As the instrument was unused for a 

long period, the memory backup 

battery was exhausted and the 

measurement results stored in the 

memory were lost.

Keep the instrument powered for at 

least 11 hours to charge the backup 

battery. After charging the instrument, 

set the date and time. (See page 3-11.)

E004

No paper in the printer

• The thermal recording paper has 

run out.

• The thermal recording paper roll 

was not properly installed.

• Install a new roll of thermal recording 

paper. (See page 4-12.)

• Make sure the thermal recording 

paper is installed correctly.

E005

Waste box is full

At the end of the measurement, the 

test strip waste counter exceeded 90.

Cancel the error notice by pressing 

 

. Empty the used test strips from the 

waste box, and press 

 .

E006

Surplus urine is full

At the end of the measurement, the 

total number of measurements since 

turning ON the instrument exceeded 

190.

Turn OFF the instrument and discard 

any surplus urine.

E007

Data not found

No data was found in the specified 

range.

Verify that the specified range for 

re-printing, re-sending, or list printing is 

appropriate.

E008

Auto start sensor error

• A test strip was placed while the 

carrying arm was moving to the 

suction ports.

• The detection section of the auto 

start sensor was soiled with urine.

• The auto start sensor is faulty.

• Remove the test strip.

• Clean the detection section.

• If you find no abnormality and the 

error recurs, contact your local 

distributor.

E009

No test strip on the 

feeder

• The feed lever is not attached.

• The incoming strip sensor is faulty.

• The sensor failed to detect a test 

strip.

• Turn OFF the instrument and open 

the maintenance cover.

• Mount the feed lever.

• If you find any obstacles or scattered 

test strips inside, remove them.

• If you find no abnormality and the 

error recurs, contact your local 

distributor.

E011

Elapse of QC deadline

The QC deadline elapsed while the 

instrument was OFF or on standby.

Perform control measurements.

Содержание AUTION ELEVEN AE-4022

Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...

Страница 2: ...e Waste Box 11 Replacing Thermal Paper 12 Warnings 13 Errors 14 15 This Quick Reference Guide is a brief summary of the operation and routine maintenance of the AUTION ELEVEN Semi Automated Urine Analyzer Read the Operating Manual before using the instrument Wear protective gloves to prevent exposure to pathogenic microbes SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 ...

Страница 3: ...System Overview System Overview ...

Страница 4: ...strip tray Carrying arm Cover open button on both sides Waste box Maintenance cover Built in printer Rails white Test strip detection window Preparation of measurement Setup Meas Check that there is no used test strips in the waste box S Select t Dip the Set the Place a strip de of the te The car analyze Set the System Overview ...

Страница 5: ...Sample Preparation Sample Preparation ...

Страница 6: ...no used test strips in the waste box Turn on the power switch at the back of the instrument Prepare a sufficient volume of sample in urine collection cup so that the entire test strip pad area can be dipped in a single motion Stir well Waste box Test strip Test strip pad area 2 3 Preparation for measurement bPress The recorded patient ID number is stored and the display returns to the Standby scree...

Страница 7: ...Sample Measurement Sample Measurement ...

Страница 8: ...umber will display alternately on screen AE 4022 2 12 1 Call up the screen aWith the Standby screen displayed press and hold for 3 seconds The Operator ID number input screen appears 2 Enter your operator ID number aEnter your operator ID number correctly using the numeric and keys bPress If your entry is valid the display returns to the Standby screen You are then permitted to perform measurement...

Страница 9: ...lection cup Set the measurement number Place a test strip over the detection area of the test strip detection window crossing the two white rails of the test strip tray The carrying arm sends the test strip into the analyzer and the measurement starts automatically Set the patient ID Do NOT wet the black mark Test strip Test strip detection window Detection up to 13 digits up to 4 digits IMPORTANT...

Страница 10: ... mark After dipping the test strip in the sample remove excess urine by letting the strip touch the edge of the collection cup Place a test strip over the detection area of the test strip detection window crossing the two white rails of the test strip tray The carrying arm sends the test strip into the analyzer and the measurement starts automatically When measuring continuously repeat steps 1 to ...

Страница 11: ...Control Measurement Control Measurement ...

Страница 12: ...trip in the sample for 2 seconds and then withdraw it IMPORTANT Dip the entire test strip pad area in the sample at once Do not allow the black mark shown in the figure to become wet If the test strip is dipped icon is displayed 2 Enter measurement number to identify control level 3 Dip entire test strip pad area in sample for two 2 seconds If buzzer setting is ON follow the timing chart below AE ...

Страница 13: ...Check Measurement Check Measurement ...

Страница 14: ... parameter See Operational mode on page 2 3 The standard setting is Auto start mode To change the operational mode when turning ON see 3 7 8 No 005 Operational mode when turning ON on page 3 8 When using the Cycle start mode press to start the check measurement of the first strip In the Auto start mode the system automatically starts the check measurement of the first strip when placement of the c...

Страница 15: ...Maintenance Maintenance ...

Страница 16: ... any dirt Carrying arm Suction port Waste box Test strip feed tray Feed lever Test strip tray Wipe the parts with cloth and fully dry them before attaching them back on This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance Before using the instrument make sure to read the Operat Clean the area circled in red below Suction port Detach the suction port Test...

Страница 17: ... feed lever by sliding the white lever in the center towards you Wash the detached carrying arm waste box test strip tray suction port test strip feed tray and feed lever with water to remove any dirt Carrying arm Suction port Waste box Test strip feed tray Feed lever Test strip tray Wipe the parts with cloth and fully dry them before attaching them back on This Quick Reference Guide is a brief su...

Страница 18: ...ve any dirt Carrying arm Suction port Waste box Test strip feed tray Feed lever Test strip tray Wipe the parts with cloth and fully dry them before attaching them back on This Quick Reference Guide is a brief summary of the measurement operation and daily maintenance Before using the instrument make sure to read the Operat ing Manual Clean the area circled in red below r Wash the detached carrying...

Страница 19: ...with alcohol then wash with water and wipe dry 3 Attach waste box 2 4 10 Items required Alcohol cloth and protective gloves 1 Discard used test strips aCheck that the Standby screen is displayed bPull out the waste box and discard the used test strips 2 Sterilize and clean the waste box aSterilize the waste box using alcohol and then wash with water to remove all traces of dirt bWipe all moisture ...

Страница 20: ...ly run out remove the paper core from the paper holder and go to step 3 2 Remove the remaining thermal recording paper aPress The thermal recording paper remaining in the printer is fed out Remove the paper by holding it with your fingertips Printer cover Take out Cut using scissors button after replacing the thermal recording paper If you do not fully feed the thermal recording paper through the ...

Страница 21: ...Warnings Warnings ...

Страница 22: ...ult is not obtained When a problem occurs during measurement the system continues measurement and prints warning messages with the measurement results The table below explains the meaning of and measures corrective action to be taken for each warning message Chapter 5 Troubleshooting 5 1 Warning messages A warning message appears when a normal measurement result was not obtained When a problem occ...

Страница 23: ...Errors Errors ...

Страница 24: ... of thermal recording paper in the paper holder orienting it correctly so that the paper feeds from the bottom Guide the trimmed end of the thermal recording paper into and through the slot that receives the recording paper The thermal recording paper is automatically wound forward and fed into the unit NOTE When the thermal recording paper is completely removed from the printer the screen message...

Страница 25: ...ty and the error recurs contact your local distributor E009 No test strip on the feeder The feed lever is not attached The incoming strip sensor is faulty The sensor failed to detect a test strip Turn OFF the instrument and open the maintenance cover Mount the feed lever If you find any obstacles or scattered test strips inside remove them If you find no abnormality and the error recurs contact your...

Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...

Отзывы: