SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Maintenance – Daily
Clean the Waste Box
Chapter 4 Maintenance
AE-4022
4-10
4.1.2 Cleaning the waste box
The waste box gets full after approximately 100 measurements. Discard the used test strips, and sterilize and clean the
waste box.
Items required: Alcohol, cloth, and protective gloves.
1 Discard used test strips
a
Check that the [Standby screen] is displayed.
b
Pull out the waste box and discard the used test strips.
2 Sterilize and clean the waste box
a
Sterilize the waste box using alcohol and then wash with water to
remove all traces of dirt.
b
Wipe all moisture off with a cloth to dry the part.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Discard used test strips in accordance with local regulations for biohazardous waste.
NOTE:
When sterilizing and cleaning the waste box, avoid the use of thinner, other organic solvents, or ultrasonic
cleaning. These sterilizing/cleaning methods may deform or discolor the test strip tray and thus make it
unserviceable for further testing.
Waste box
1. Remove waste box and
discard used test strips.
2. Sterilize and clean waste box.
• Sterilize with alcohol then
wash with water and wipe dry.
3. Attach waste box.
Chapter 4 Maintenance
AE-4022
4-10
4.1.2 Cleaning the waste box
The waste box gets full after approximately 100 measurements. Discard the used test strips, and sterilize and clean the
waste box.
Items required: Alcohol, cloth, and protective gloves.
1 Discard used test strips
a
Check that the [Standby screen] is displayed.
b
Pull out the waste box and discard the used test strips.
2 Sterilize and clean the waste box
a
Sterilize the waste box using alcohol and then wash with water to
remove all traces of dirt.
b
Wipe all moisture off with a cloth to dry the part.
Wear protective gloves to prevent exposure to pathogenic microorganisms.
Discard used test strips in accordance with local regulations for biohazardous waste.
NOTE:
When sterilizing and cleaning the waste box, avoid the use of thinner, other organic solvents, or ultrasonic
cleaning. These sterilizing/cleaning methods may deform or discolor the test strip tray and thus make it
unserviceable for further testing.
Waste box
4.1 Daily maintenance
AE-4022
4-11
3 Attach the waste box.
a
Attach the waste box to the unit.
NOTE:
When attaching the waste box to the unit,
do not
spread
tissue paper (Kleenex) or anything else in the bottom of the
box. Doing so may cause a problem later when disposing of
waste.
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...