2
SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Sample Preparation
1.
Turn on power switch at back of instrument.
2. After a 20 second warm up the standby screen will appear.
3.
Prepare suffi cient volume of sample in urine collection cup so entire test strip pad
area can be dipped in a single movement.
4. Mix well.
Daily Maintenance
Power switch
Turn off the instrument.
Waste box
Discard used test strips.
140820YH
Wear protective gloves to prevent
exposure to pathogenic microbes.
Discard used test strips and
protective gloves separately from
general waste according to local
regulations on biohazardous waste.
Detach the carrying arm.
Open the maintenance cover.
Feed lever
Lever
Detach the feed lever by sliding the white lever
in the center towards you.
Wash the detached carrying arm, waste box,
test strip tray, suction port, test strip feed tray
and feed lever with water to remove any dirt.
Carrying arm
Suction port
Waste box
Test strip feed tray
Feed lever
Test strip tray
Wipe the parts with cloth and fully dry them before
attaching them back on.
This Quick Reference Guide is a brief
summary of the measurement operation
and daily maintenance. Before using the
instrument, make sure to read the Operat-
ing Manual.
Clean the area circled in red below.
Test strip feed tray
Detach the test strip feed tray.
Suction port
Test strip tray
Detach the suction port.
Test strip tray
Detach the test strip tray.
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with
buttons
Select
[ Select test strip type ]
[ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
2.3 Preparation for measurement
AE-4022
2-15
b
Press .
• The recorded patient ID number is stored and the display returns
to the [Standby screen].
• Once the patient ID number is set, the “measurement
test strip type” and “patient ID number” are displayed alternately
on the [Standby screen].
To enter a patient ID number using a barcode
1 Call up the screen
a
With the [Standby screen] displayed, press
.
• The [Patient ID number input screen] appears.
2 Enter patient ID number by reading its barcode
a
Use a hand-held barcode reader to scan the barcode on the label,
which will enter the ID number when the [Patient ID number
input screen] is displayed.
Displayed alternately
IMPORTANT:
Do not
centrifuge samples.
Centrifugal force may cause undesirable sedimentation of blood cells and/or other
sample contents, thus causing several measurement items to have incorrect results.
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...