SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Check Measurement
1. Clean the feeder and waste box to prevent check strips from becoming soiled.
2. Change measurement mode.
• From the [standby screen] press
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with
buttons
Select
[ Select test strip type ]
[ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
or
Measurement
Name of each part
Display
Operator panel
Test strip tray
Carrying arm
Cover open button
(on both sides)
Waste box
Maintenance cover
Built-in printer
Rails (white)
Test strip detection
window
Preparation of measurement
Setup (test strip, measurement No., patient ID)
Measurement
Check that there is no used test strips in the waste box.
with
buttons
Select
[ Select test strip type ]
[ Menu screen ]
Select the type of test strip
Dip the entire test strip pad area in the sample for two seconds.
IMPORTANT
Do NOT wet the black mark.
After dipping the test strip in the sample,
remove excess urine by letting the strip
touch the edge of the collection cup.
Set the measurement number
Place a test strip over the detection area of the test
strip detection window, crossing the two white rails
of the test strip tray.
The carrying arm sends the test strip into the
analyzer, and the measurement starts automatically.
Set the patient ID
When measuring continuously, repeat steps 1 to 3 .
Set measurement numbers and patient IDs as necessary.
When the buzzer setting is ON, the beeper sounds
for two seconds to signal the timing for dipping the
next test strip.
Dip the test strip while you hear the beep.
Turn off the instrument after completing all the measurement procedures.
Turn on the power switch at the back of the instrument.
Prepare a sufficient volume of sample in urine collection cup so
that the entire test strip pad area can be dipped in a single
motion. Stir well.
Waste box
Test strip
Test strip
pad area
Do NOT wet the
black mark.
Test strip
Rails (white)
Test strip detection window
Detection
area
(up to 13 digits)
,
(up to 4 digits)
until the
2-27
AE-4020
Chapter 2 Measurement Operation
3
Prepare the check strips
Check strip (white)
Check strip (gray)
Reflectance label
Prepare the check strips. The check strip bottle has
a reflectance label on it, which must be used when
evaluating reflectance after the check measurement.
Keep this labeled bottle in an easily accessible
place.
4
Start the check measurement
Place a white check strip on the test strip tray.
REFERENCE:
Place the check strip on the test strip tray with the black mark facing up.
NOTE:
Place a test strip in the detection
area of the test strip detection
window as shown in the right figure.
If the system does not detect the test
strip in the auto start mode, remove
and replace it in the center of the test
strip detection widow. At this time,
the test strip must cross both of the
white rails in the detection area. If the
test strip is not placed in the correct
position, it may not be transported
properly, leading to a strip jam or
inaccurate measurement result.
Rails (white)
Test strip
detection window
Detection
area
Pay attention to the carrying arm when it moves, so as not to
have your fingers caught or pinched.
IMPORTANT:
Different procedures at this stage are required, depending on the
setting of the “operational mode when turning ON” parameter. (See
“Operational mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the operational
mode when turning ON, see “3-7-8. No.005: Operational mode when
turning ON” on page 3-8.
● When using the “Cycle start” mode, press to
start the check measurement of the first strip.
● In the “Auto start” mode, the system automatically
starts the check measurement of the first strip
when placement of the check strip is detected.
icon is displayed.
3. Place white check strip on test strip tray and perform measurement.
4. Place gray check strip on test strip tray and perform measurement.
5. Remove check strips from waste box and return to check strip bottle.
6. Evaluate results
• Compare the refl ectance values on the test result printout to the ranges on the refl ectance label on the check strip bottle.
• If any of the measurement results are outside of the specfi ed ranges, repeat check measurement using the other set of
check strips.
• If any of the measurement results are still outside the ranges, contact ARKRAY Technical Support at 1.855.646.3108.
2.4 Measurement operation
AE-4022
2-33
b
Press
or
to switch to check measurement mode.
• On the [Standby screen], the
icon is displayed, and the
measurement mode is switched to the Check measurement mode.
3 Prepare the check strips
a
Prepare the check strips. The check strip bottle has a reflectance
label on it, which must be used when evaluating reflectance after
the check measurement. Keep this labeled bottle in an easily
accessible place.
4 Start the check measurement
a
Place a white check strip on the test strip tray.
Reflectance label
Check strip (gray)
Check strip (white)
REFERENCE:
Place the check strip on the test strip tray with the black mark facing up.
Detection
area
Test strip
detection window
Rails (white)
NOTE:
Place a test strip in the detection area of the test strip
detection window as shown in the right figure. If the system
does not detect the test strip in the auto start mode, remove
and replace it in the center of the test strip detection
window. At this time, the test strip must cross both of the
white rails in the detection area. If the test strip is not placed
in the correct position, it may not be transported properly,
leading to a strip jam or inaccurate measurement result.
Pay attention to the carrying arm when it moves, so as not to have your fingers caught or
pinched.
IMPORTANT:
Different procedures at this stage are required, depending on the setting of the “operational mode when
turning ON” parameter. (See “
Operational mode” on page 2-3.)
The standard setting is “Auto start” mode. To change the operational mode when turning ON, see “3.7.8.
No.005: Operational mode when turning ON” on page 3-27.
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...