SEMI-AUTOMATED URINE ANALYZER
AUTION ELEVEN
™
AE-4022
Warnings
Message Error
Possible cause
Measures (Corrective Action)
W001
Drift
The ambient light level around the instrument changed
rapidly, due to a photofl ash or other reasons.
Maintain a constant level of illumination
around the instrument during measurement.
W002
Abnormally high
refl ectance
An abnormal sample (e.g. a sample containing
medicine) was measured.
Check if the tested sample is abnormal or
not. If the sample is normal, test it again.
W003
Incorrect test strip
position
• The test strip was displaced from its correct position
due to vibration.
• The test strips were incorrectly placed.
• Do not expose the instrument to any
vibration during measurement.
• Place the test strips correctly on the test
strip tray.
W004
No dipping
The entire pad section of the test strip was not dipped.
Dip a new test strip in the sample correctly
and test again.
W005
Wrong test strip
A test strip other than that specifi ed was used.
Only use ARKRAY’s test strips designed for
the AUTION ELEVEN.
W006
Transport error
The test strip may not have been transported properly.
• Clean the test strip tray and suction ports.
• Attach the suction ports properly.
• Clean the incoming strip sensor window.
W007
Number of registered
operator ID numbers
150 operator ID numbers have been registered.
Delete unnecessary operator ID numbers
before registering new ones.
W008
Elapse of QC deadline The QC deadline has elapsed.
Perform control measurements.
A warning message appears when a normal measurement result is not obtained.
When a problem occurs during measurement, the system continues
measurement and prints warning messages with the measurement results.
The table below explains the meaning of and measures (corrective action) to be
taken for each warning message.
Chapter 5 Troubleshooting
AE-4022
5-2
5.1
Warning messages
A warning message appears when a normal measurement result was not obtained.
When a problem occurs during measurement, the system continues measurement and prints warning messages with
the measurement results.
The table below explains the meaning of and measures to be taken for each warning number (message).
Message
Problem
Possible cause
Measures
W001
Drift
The ambient light level around the
instrument changed rapidly, due to a
photoflash or other reasons.
Maintain a constant level of
illumination around the instrument
during measurement.
W002
Abnormally high
reflectance
An abnormal sample (e.g. a sample
containing medicine) was measured.
Check if the tested sample is
abnormal or not. If the sample is
normal, test it again.
W003
Incorrect test strip
position
• The test strip was displaced from its
correct position due to vibration.
• The test strips were incorrectly
placed.
•
Do not
expose the instrument to
any vibration during measurement.
• Place the test strips correctly on the
test strip tray. (See page 2-20.)
W004
No dipping
The entire pad section of the test
strip was not dipped.
Dip a new test strip in the sample
correctly and test again.
W005
Wrong test strip
A test strip other than that specified
was used.
Only use ARKRAY’s test strips
designed for the AUTION ELEVEN.
W006
Transport error
The test strip may not have been
transported properly.
• Clean the test strip tray and suction
ports.
• Attach the suction ports properly.
• Clean the incoming strip sensor
window.
W007
Number of registered
operator ID numbers
150 operator ID numbers have been
registered.
Delete unnecessary operator ID
numbers before registering new
ones.
W008
Elapse of QC deadline
The QC deadline has elapsed.
Perform control measurements.
Warning number
Содержание AUTION ELEVEN AE-4022
Страница 1: ...SEMI AUTOMATED URINE ANALYZER AUTION ELEVEN AE 4022 Quick Reference Guide ...
Страница 3: ...System Overview System Overview ...
Страница 5: ...Sample Preparation Sample Preparation ...
Страница 7: ...Sample Measurement Sample Measurement ...
Страница 11: ...Control Measurement Control Measurement ...
Страница 13: ...Check Measurement Check Measurement ...
Страница 15: ...Maintenance Maintenance ...
Страница 21: ...Warnings Warnings ...
Страница 23: ...Errors Errors ...
Страница 26: ...Minneapolis MN 55439 877 538 8872 corelab arkrayusa com www arkrayusa com 5266 00 Rev 1 17 ...