32
Service Procedures
SP7 - Scale Installation on a Non-Scale
SP7 - Scale Installation on a
Non-Scale Unit
The installation of a scale on a non-scale unit
(KMX
Q
X
Q
) is similar to the scale
replacement procedure, but with some extra
steps for the certification.
Ordering Kits
Prior to installing the scale, the correct kit
must be ordered. The service kit needed
depends on the Maxi Move model and the
user’s location:
Verified Scale (CE Country)
Follow the instructions in the
Maxi Move
Maintenance and Repair Manual
001.25075.EN
under chapter 8 “Jib, Scale
and T-bar” for the mechanical installation
procedure for the scale.
Once the scale has been installed:
1) With the marker supplied in the kit, write
the serial number of the Maxi Move on
the scale marking label (in the box next to
the letters “SN:”, as shown in Fig.13).
Fig.13
Apply the scale marking label on the control
box, above the handset connection (see ,
callout 1). Remove the clear plastic laminate
backing and carefully apply it over the scale
marking label.
2) Apply the scale handset label on the
handset (see , callout 2).
3) Apply the scale logo to the Maxi Move
top cover (see , callout 3).
4) Replace the model label with the one
supplied in the upgrade kit. Note: In kit
700-25335, two different label models are
supplied (for KMCSXN and KMCLXN).
Use the label corresponding to the model
of the unit being upgraded and discard the
M
AXI
M
OVE
MODEL
S
ERVICE
KIT
NUMBER
S
ERVICE
KIT
DESCRIPTION
KMC
Q
XN in
a non-CE
country
700-25340
Maxi Move stan-
dard jib (non-CE)
scale kit upgrade
KMC
Q
XN in
a CE country
700-25335
Maxi Move stan-
dard jib (CE) scale
kit upgrade
KMC
Q
XE in a
CE country
700-25350
Maxi Move
extended jib (CE)
scale kit upgrade
Note: You will need to provide the serial
number of the unit being upgraded
when ordering the required kit.
Note: The main board will automatically
detect the addition of the scale.
However, the DIP switch settings must
be set according to those found in the
Maxi Move Maintenance and Repair
Manual, under service procedure 10.
5
(
9
$UMR+XQWOHLJK$%
+DQV0LFKHOVHQVJDWDQ
0DOP|
6:('(1
HJ
NJ
92,'
92,'
NJ
0LQ
&&
0D[
&DOLEUDWLRQFRXQWHU
*UDYLW\FRXQWHU
75$1*(
6&$/(
&(57,),&$7(7
&DW
Using the marker
provided, write the
Maxi Move’s serial
number within the box.
Note: Use ONLY the permanent marker
supplied in the kit to ensure that what is
written down will not be later erased by
the use of cleaning products on the
unit.