Arm and Leg Garments
EN
WARNINGS
•
Do not apply the garment to the limb unless
it is partially zipped, as you may damage the
garment zip.
• Do not undo the zips or attempt to remove the
garments during the therapy session or you
may damage the zips. Make sure the therapy
session has stopped and the garments have
deflated before you remove the garments.
•
Do not stand or walk with the garments fitted.
•
In the event of a power failure or fault whereby
the garment remains inflated, disconnect the
tubeset(s) in order to deflate the garment(s)
and then remove the garment(s) from the
limb(s).
Disinfecting the Garment
C
1000ppm
NaOCl
NaDCC
After cleaning, wipe complete garment
over using 70% isopropyl alcohol wipe or
a chlorine-releasing agent at 1000 p.p.m.
available chlorine.
Do not use phenol or phenol-derivative
disinfectant.
Rinse with clean water to remove any residue.
Air dry thoroughly.
Cleaning the Garment
Wipe down using a neutral detergent or
soap powder at 51°C (120°F).
Do not iron
Do not dry
clean
Do not tumble
dry
Do not machine wash.
Air dry thoroughly.
Cleaning the Tubeset
Use a soft brush. Air dry only.
Do not immerse in water. Do not machine wash.
For use with
Huntleigh Hydroven
®
3
pump only.
Refer to the
Hydroven 3
User Manual for complete
information on using the pump and garments.
100% Latex Free.
Applying the Garment
1.
Check the type of garment - Hydroven 1 (single
chamber) or Hydroven 3 (three chamber) - and the
length of the garment required.
2. Undo the zip on the garment.
3. If a garment insert is fitted to the garment, fully
fasten one of the zips between the garment and
insert, leaving the other unfastened.
4.
Before applying the garment (and insert, if fitted)
to the limb, zip up the first 150 mm (6”) of the
unfastened garment zip. Put the garment (and
insert) onto the limb and fully fasten the zip. Make
sure that the quick air release bung is secured.
Quick air release bung
Hydroven 3 Leg Garment
Tubing to pump
Hydroven 3 Arm Garment
5. Make sure the patient is in a comfortable position
with the limb supported or elevated as necessary.
6.
Check that the insert piece connecting tube is not
kinked and is attached to the garment using the
lower outlet bung as shown below.
Arm Garment with
insert fitted
Leg Garment
with insert
fitted
7. Attach the garment tubing to the pump ensuring a
“click” is heard from each snap-lock connector.
8.
If only one garment is to be used, attach the
garment to either port on the pump. The system
will automatically identify that only one garment is
to be used.
9. Switch on the pump and adjust the pressure
control accordingly (refer to the pump user
manual).
To be used only under the direction of a
physician.
Before fitting the garment ensure all quick air
release bungs are closed, as this will effect the
efficiency of the garment.
UK marking indicating conformity with UK Medical Devices
Regulations 2002 (SI 2002 No 618, as amended)
Figures indicate UK Approval Body supervision.
UK Responsible Person & UK Importer:
Arjo (UK) Ltd., ArjoHuntleigh House, Houghton Regis. LU5 5XF
Is the appointed UK Responsible Person as defined in UK Medical Devices
Regulations 2002 (SI 2002 No 618, as amended).
For Northern Ireland (NI) CE marking will still apply until further amendment to
applicable regulations.
This section is only applicable to United Kingdom (UK) market when UK
marking is applied to the Arjo medical device labelling.
UK Symbol:
1001071-2