Backup battery
The backup battery allows operation of the bed for short periods when it is
disconnected from the electricity supply or in emergency situations when the
electricity supply is not available.
The battery’s charge level is indicated as follows:
If an intermittent warning tone (beep-beep-beep) sounds when operating
the bed, the battery is between 75% and 100% charged.
In this condition all bed functions remain operational.
If a continuous warning tone sounds when operating the bed, the battery is
between 10% and 75% charged.
In this condition, all bed functions remain operational.
If the ACP battery indicator lights red, the battery is less than 10% charged.
In this condition, all functions are locked.
CAUTION
To ensure the battery is kept fully charged and prevent damage to the battery, the bed
should be connected to the electricity supply at all times during normal use.
The battery is intended for short term use only. Its life will be reduced if it is used to
power the bed for long periods.