41
Initialisation
After connecting the bed to the electricity supply, the weighing system must
be initialised as follows:
1. Place the mattress, bed linen and all necessary accessories on
the bed. The patient must not be on the bed at this point.
2. Press the
Zero
button once.
3. After a few seconds the display will read
.
4. The patient can now get onto the bed
5. The bed cannot be re-zeroed once a patient is on the bed. Use
the Auto Compensation feature if adding or removing items from
the bed such as linen or accessories.
WARNING
The weighing system is intended to provide reference data only. The weighing system
is not intended to provide readings upon which medication dosage decisions are
made.
CAUTION
The weighing system MUST be zeroed every time a new patient is to be placed on the
bed.
CAUTION
The weighing system MUST be zeroed every time a mattress is changed.
CAUTION
The weighing system cannot be zeroed if a mattress or accessories weighing more
than 35kg are fitted. To ensure the bed can be zeroed correctly, only use Arjo
approved mattresses.
Содержание E9X
Страница 1: ...Enterprise 9000X E9X INSTRUCTIONS FOR USE 746 591 EN_13 06 2019...
Страница 65: ...65 Intentionally left blank...
Страница 66: ...66 Intentionally left blank...