42
Weighing
To calculate and display the patient’s weight:
1. With the patient lying centrally on the bed, press the
Weigh
button once.
2. The display will show a moving circular pattern until a stable
weight reading is obtained.
3. The patient’s weight will be displayed for ten seconds then the
display will go blank.
The patient’s weight is usually displayed rounded to the nearest 500g; this
can be temporarily changed to 100g by pressing the
Select Weight
Display
button. The display will revert to 500g next time the
Weigh
button
is pressed.
Auto
Compensation
The Auto Compensation facility allows weight (up to a maximum of 100kg)
to be added to or removed from the bed, without affecting the indicated
patient weight.
1. With the patient lying on the bed, press the
Auto Compensation
button once.
2. The display will show a moving circular pattern until a stable
weight reading is obtained.
3. The display will show
to indicate that the system is in Auto
Compensation mode.
4. Add or remove accessories, bed linen, pillows, etc. as required.
5. Press the
Auto Compensation
button again.
6. The display will show a moving circular pattern for a few seconds
then revert to showing the patient’s weight..
NOTE
The 100g weight display should only be used for comparative readings.
CAUTION
The weighing system requires periodic calibration to ensure accuracy. See
“Preventive maintenance” on page 49.