CLEANING AND DISINFECTION
Accessories needed for cleaning/ disinfection
• Protective
glasses
• Protective
gloves
• Spray bottle with cleaning solution
• Spray bottle with disinfectant solution
• Spray bottle with water
• Cloths
Clean and disinfect
(26 steps)
Always follow these steps for proper cleaning and disinfection after each patient.
Prepare the pump
1. Disconnect the pump from the mattress.
2. Select Standby on the pump unit. Disconnect the pump from the power source
.
Clean the pump
3. Put on protective glasses and gloves.
4. Spray the cleaning solution on a clean cloth.
5. Wipe all areas of the pump to remove any deposits or visible dirt.
6. Clean all areas with residual dirt with the cloth as needed.
7. Use a new cloth soaked in water to wipe o
ff
all traces of cleaning solution.
8. Use a dry cloth to remove any excess moisture from the pump.
Disinfect the pump
9. Spray disinfectant solution on a clean cloth and wipe all areas of the pump.
10. Allow time for disinfection according to the instructions provided by the disinfectant manufacturer.
11. Use a new cloth soaked in water to wipe o
ff
all traces of disinfectant solution from the pump.
12. Use a dry cloth to remove any excess moisture from the pump.
13. Allow the pump to air dry.
Instructions continue on the next page