47
ELECTROMAGNETIC
COMPATIBILITY
Electromagnetic compatibility
The hybrid mattress system is intended for use in the electromagnetic environment speci
fi
ed below. The
customer or the user of the hybrid mattress system should make sure that it is used in such an environment.
Electromagnetic Compatibility (EMC)
This product complies with the requirements of applicable EMC Standards. Medical electrical equipment
needs special precautions regarding EMC and needs to be installed in accordance with the following
instructions:
• The use of accessories not speci
fi
ed by the manufacturer may result in decreased immunity of the product,
or increased emissions from the product. This would a
ff
ect the products performance.
• Portable and mobile radio frequency (RF) communications equipment (e.g. mobile/cell phones) can a
ff
ect
medical electrical equipment.
• If this equipment needs to be used adjacent to other electrical equipment, normal operation must be
checked before use.
For detailed EMC information contact Arjo service personnel.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSION
The pump is intended for use in the electromagnetic environment speci
fi
ed below. The customer or the user of the
pump should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
Group 1
The pump uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
Class B
This pump is suitable for use in all establishments including
domestic and those directly connected to the public low voltage
power supply network.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fl
uctuations/
fl
icker emissions
IEC 61000-3-3
Complies
WARNING
Stacking or placing other electrical equipment next to this device is not recommended, it can
interfere with the equipment’s operation and safety. Portable and mobile radio frequency (RF)
communications equipment can interfere with this device operation and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby device. It may
be necessary to take action, such as reorienting, relocating the device or shielding the location.