INTENDED USE
7
INTENDED USE
Contraindications
In powered active alternating mode with the pump,- Do not use the hybrid mattress system with
patients with an unstable cervical, thoracic and/or lumbar fracture, cervical traction, and skeletal
traction. For any other conditions that may be complicated by a moving surface, do not use the hybrid
mattress system.
In the non-powered reactive mode without the pump, - It may be possible to use the surface for
patients with unstable cervical, thoracic and/or lumbar fracture, cervical traction, and skeletal traction
if assessed by a healthcare professional as suitable to do so. Ongoing assessment and continuous
monitoring of the patient is advised.
Patient assessment
Facilities should establish regular assessment routines. Caregivers should assess each patient before
using the product. The patient weight must not exceed:
• 250 kg (550 lb) for the Standard mattress
• 454 kg (1000 lb) for the Plus (bariatric) mattress
• 250 kg (550 lb) for the Stretcher (ST) mattress
• 250 kg (550 lb) for the Seat cushion
If the patient does not meet these criteria an alternative medical device/system shall be used.
Expected service life
The expected service life of the AtmosAir Velaris system elements is:
•
Mattress – 5 Years
•
Seat Cushion – 5 Years
•
Pump – 7 Years
The expected service life of this device is subject to preventive maintenance being carried out in
accordance with the instructions for care and maintenance found in this Instructions for Use.