49
ELECTROMAGNETIC
COMPATIBILITY
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF
COMMUNICATION EQUIPMENT AND THE PUMP
The pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the pump can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the pump as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter - W
Separation distance according to frequency of transmitter - m
150 kHz to 80 MHz
d = 1.2
√
P
80 MHz to 800 MHz
d = 1.2
√
P
800 MHz to 2.7 gHz
d = 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
2.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (w) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note: These guidelines
may not apply in all situations. Electromagnetic propagation is a
ff
ected by absorption and re
fl
ection from structures,
objects and people.