AT680 Leakage Current/IR Meter user’s Manual
50
Insulation Resistance
:
Accuracy
<1G:
1%
1G:
3%
10G:
5%
M=10
6
,G=10
9
Range
voltage
1 2 3 4 5 6 7
1V
--- ---
5k-50k
50k-500k
500k-5M
5M-50M
50M-500M
10V
---
5k-50k 50k-500k 500k-5M 5M-50M 50M-500M 500M-5G
25V
1.25k-12.5k 12.5k-125k 125k-1.25M 1.25M-12.5M
12.5M-125M 125M-1.25G 1.25G-12.5G
50V
2.5k-25k 25k-250k 250k-2.5M 2.5M-25M 25M-250M 250M-2.5G 2.5G-25G
75V
3.75k-37.5k 37.5k-375k 375k-3.75M 3.75M-37.5M
37.5M-375M 375M-3.75G 3.75G-37.5G
100V
5k-50k 50k-500k 500k-5M 5M-50M 50M-500M 500M-5G 5G-50G
125V
6.25k-62.5k 62.5k-625 625k-6.25M 6.25M-62.5M
62.5M-625M 625M-6.25G 625M-62.5G
250V
12.5k-125k 125k-1.25M 1.25M-12.5M
12.5M-125M 125M-1.25G 1.25G-12.5G 12.5G-125G
500V
25k-250k 250k-2.5M 2.5M-25M 25M-250M 250M-2.5G 2.5G-25G 25G-250G
650V
32.5k-325k 325k-3.25M 3.25M-32.5M
32.5M-325M 325M-3.25G 3.25G-32.5G 32.5G-325G
Display:
Vacuum-Fluorescent-Display
(4-Colors
VFD)
Size:
98x55mm
Output Voltage:
-1.0VDC ~ -650VDC
Voltage Accuracy
:
0.5%
Parameter
:
Insulation Resistance, Leakage Current, Peak and Comparator Result.
Measurement Range
:
Leakage Current
:
1nA~20mA
Resistance
:
1k
~ 325G
Basic Accuracy
:
(@Slow Rate)
:
Current
:
0.5%
Resistance
:
<10G:
3%
10G:
5%
100G:
10%
Maximum counts
:
Slow:
3 readings/second with 9,999 counts
Medium:
25 readings/second with 1,999 counts
Fast:
55 readings/second with 1,999 counts
Max Charge Current: 200mA
20mA
Rate
:
Fast
:
55 readings/
,
s Medium
:
25 readings/
,
s Slow
:
3 readings/s
Charge
Time:
999.9s
0.1s
resolution
Accuracy:
0.5%
Sample
Time:
999.9s
0.1s
resolution
Accuracy:
0.5%
Trigger:
Internal,
Manual
(Remote) and External Trigger
Range:
Auto
and
Manual
Correction:
Open
Clear
Zero
Comparator:
30 sets of record. Display and Output GD/GD.
Beep: GD,
NG
and
OFF
Interfaces: Built-in
Handler
Interface.
Built-in RS232C Interface.
Program Language:
SCPI
Environmental
:
18
℃
~28
℃
stated
accuracy
65% RH
Operating
:
10 ~40
℃
℃
stated accuracy 10~80% RH
Storage
:
0 ~50
℃
℃
stated accuracy 10~90% RH