AT680 Leakage Current/IR Meter user’s Manual
40
Parameters
{<integer>,MIN,MAX}
where
,
<interger>
Range no from 1 to 7.
MIN
=1
MAX
=7
For example
:
Tx>
func:rang 5
//set range to 5
Tx>
func:rang min
//set range to 1
Tx>
func:rang max
//set range to 7
Query Syntax
FUNCtion:RANGe?
Query Response
<
integer
>
Range no from 1 to 7.
For example
:
Tx>
func:rang?
Rx>
6
Constraints
none
:RANGe:AUTO
The
:RANGe:AUTO
command sets the auto range to ON or OFF.
Command Syntax
FUNCtion:RANGe:AUTO {ON,OFF,1,0}
Parameters
{ON,OFF,1,0}
where,
1
:
=ON
,
ASCII(decimal 49)
,
0
:
=OFF
,
ASCII(decimal 48)
For example
:
Tx>
func:rang:auto off
//The auto range will set to manual.
Query Syntax
FUNCtion:RANGe:AUTO?
Query Response
{on,off}
For example
:
Tx>
func:rang:auto?
Rx>
off
Constraints
none
7.7 VOLTage Subsystem
The
VOLTage
subsystem sets test voltage.
Figure7-3
VOLTage
Command Tree
Command Syntax
VOLTage <float>
Parameters
<float>
1.0~650
For example
:
Tx>
VOLT 10.2
Tx>
VOLT
500
Query Syntax
VOLTage?
Query Response
<float>
1.0~650.0
VOLTage
<float>