Handler
31
5.3 Timing Chart
TRIG
EOC
GD/NG
t2
t1
t3
t4
有有
测测测测
前前前前
t6
下下下下下下
t5
1: Vali d
t7
CHARG:
0: Vali d
State:
放放前前
t8
t9
DISCH:
0: Vali d
t10
新
COMP.4-0
0
0
新
0
t11
充放
/
测测前前
t12
Figure 5-4 Timing Chart
Description
Time
(
App.
)
Min. Typical Max.
t1
One Measurement Circle
Fast - - 17ms
Med - - 67ms
Slow - - 260ms
t2 Trigger
pulse
width
1ms
-
-
t3 Trig
Delay
Time
-
1ms
t4
AD Time
(
EOC[BUSY]
)
Fast - -
12.2ms
Med - -
62.6ms
Slow - - 256ms
t5 Print
Result
Time
4ms 4.4ms 4.8ms
t6
Handler Out to EOC Time
-
10
s
-
t7 Next
Start
Time
-
10
s
-
t8
CHARG Signal Hold Time
50ms
-
110ms