AT680 Leakage Current/IR Meter user’s Manual
42
:RECord
The
:RECord
command sets Record number.
Command Syntax
COMParator:RECord <integer>
Parameters
<integer>
1-30
For example
:
Tx>
COMP:REC 2
//sets current record to 2
Query Syntax
COMParator:RECord?
Query Response
<integer>
1-30
For example
:
Tx>
COMP:REC?
Rx>
2
Constraints
Available under DISCH state
:RESistance
The
:RESistance
command sets resistance limit value.
Command Syntax
COMParator:RESistance <float>
Parameters
<float>
fix float or floating point
,
0-99999G
For example
:
Tx>
COMP:RES 100G
//100G
Ω
Query Syntax
COMParator:RESistance?
Query Response
<float>
floating point
For example
:
Tx>
COMP:RES?
Rx>
1.08
Constraints
Available under DISCH state
:CURRent
The
:CURRent
command sets current limit value.
Command Syntax
COMParator:CURRent <float>
Parameters
<float>
fix float or floating point,
0-99999m
For example
:
Tx>
COMP:CURR 1m
//1mA
Query Syntax
COMParator:CURRent?
Query Response
<float>
floating point
For example
:
Tx>
COMP:CURR?
Rx>
1.000000e-06
Constraints
Available under DISCH state
:BEEP
The
:BEEP
command set beep to ON or OFF.
Command Syntax
COMParator:BEEP {on(1),off(0)}
Parameters
{on(1),off(0)}