AT680 Leakage Current/IR Meter user’s Manual
26
4
Comparator
This chapter provides information about comparator
:
Inputs limit reference
value.
Setup beep feature.
4.1 Input Limit reference value
1
.
Under DISCHARGE STATE, Press Refer key, a
。
cursor flashed
2
.
Press Digits to input value.
3
.
Press Shift + n/
/m/k/M/G to key in unit. The
value saved.
Press Esc to cancel and exit to DISCHARGE
STATE.
NOTE
:
The Resistance’s limit reference value is UPPER limit. Rx
Refer, PASS (GD).
The Leakage Current’s limit reference value is LOWER limit. Ix
Refer, PASS.
4.2 Set Beep Feature:
4.2.1 Turning on/off the beep
:
Press Beep to turn on/off the beep.
The beep annunciator shows you the beep on/off state.
4.2.2 Set Beep
:
.
1 Press Shift Beeper key to enter beep set window.
.
2 Press< > key or Turn Knob to choose following items:
GD
Beep
while
pass.
NG
Beep while fail.
.
3 Press Enter key to exit to DISCHARGE STATE and the setting saved.