AT680 Leakage Current/IR Meter user’s Manual
44
:CHARge
Command Syntax
STATe:CHARge
Parameters
none
For example
:
Tx>
STAT
?
Rx>
discharge
Tx>
STAT:CHAR
Tx>
STAT
?
Rx>
charge
Tx>
STAT:CHAR
Query Syntax
none
Query Response
none
Constraints
none
:DISCharge
Command Syntax
STATe:DISCharge
Parameters
none
Query Syntax
none
Constraints
none
7.11 TIMEr Subsystem
The TIMEr subsystem sets charge timer and sample timer.
Figure 7-7
TIMEr
Command Tree
:CHARge
The
:CHARge
command sets charge timer value.
Command Syntax
TIMEr:CHARge <float>
Parameters
<float>
fix float or floating point,0-999.9
For example
:
Tx>
TIME:CHAR 100.1
//100.1s
Rx>
TIME
0
//The charge timer sets to OFF
Query Syntax
TIMEr?
<float>
TIMEr
:CHARge
:SAMPle
<float>
<float>