AT680 Leakage Current/IR Meter user’s Manual
12
Insulation Resistance Basic Accuracy: <10G:3%
,
10G: 5%
,
100G: 10%
Max Charge Current: 200mA
20mA
Automatic Test with 7 Ranges and Manual
Test Rate
Slow: 3 readings/second with 9,999 counts
Medium: 25 readings/second with 1,999 counts
Fast:
55 readings/second with 1,999 counts
Built-in 2 timers
Charge Timer: 0s~999.9s
Sample Timer: 0s~999.9s
Trigger mode:
Internal Trig, Manual (Remote) Trig and External (Handler) Trig.
2.4 Feature overview
High brightness VFD
window size: 98mm×58mm
Correction (Zeroing) Function
Zero out test lead and fixture measurement errors.
Built-in Comparator (Sorting)
Thirty sets of Record can be used to store user’s data. Display on VFD Screen and/or
Output to Handler.
Beep and VFD Brightness can be Adjusted
Setup GD or NG Beep and adjust VFD Brightness.
Interfaces
1. Handler interface: GD/NG Output, Trig Signal Input and EOC (Busy) Output.
2. RS232C interface: SCPI Compatibility, ASCII Transmission.