![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315099.webp)
C-12
CALF—Calibration Functions
ST: In case of the ST series, do not change the settings. If changed, weighing will not be performed properly.
Function #
Function name
Range
Default
CALF-01 ST
Weighing unit
2
0 None
1 g
2 kg
3 t
CALF-02 ST
Decimal point position
0
0 None
1 2 3 4 5
1
10
1
1 2 3 4.5
2
10
2
1 2 3.4 5
3
10
3
1 2.3 4 5
4
10
4
1.2 3 4 5
CALF-03 ST
Minimum division
1, 2, 5, 10, 20, 50 (D)
1
Minimum division (D) for the weight. Input 1, 2, 5, 10, 20, or 50(decimal point ignored).
CALF-04 ST
Capacity
0 – 800000
Decimal point ignored
10000
Capacity of the weighing. Weighing is allowed up to this s 9D(9 divisions). A weight beyond
this limit is an overflow and not displayed.
CALF-05
Zero range
0 - 30
2
A range of accepting "zero" from the [ZERO] key or Control I/O. Represented in terms of percent (%)
with respect to the weighing capacity, centering to the zero calibrated point in calibration. If this is set
to 2, "zero" can be accepted within a range of the zero calibrated point +/- 2 %.
Содержание Explosion Protected Weighing Indicator AD-4403-FP ST Series
Страница 22: ...2 7 Blank page...
Страница 36: ...6 3 Blank page...
Страница 38: ...6 5...
Страница 40: ...6 7...
Страница 53: ...6 20 Blank page...
Страница 73: ...8 18 Blank page...
Страница 83: ...Dimensions OP 20 Stand A 5...
Страница 85: ...A 7 Blank page...
Страница 87: ...B 2 Blank page...
Страница 105: ...Fig Apx D 02 Gravity map WMF D 2...
Страница 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...