![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315056.webp)
8-1
8. INTERFACE
8-1 Control Input / Output
Section related:
{Appendix C: Function List—InF: Control Input Functions}
{Appendix C: Function List—outF: Control Output Functions}
The Control Input / Output interface, consists of 6 inputs and 6 outputs, is intended for
inputting/outputting weighing control signals to an external unit.
The capability of each input/output terminal can be selected at its function setting, except COM
(common).
Settings
Control Input
*
A signal is accepted at the timing when the level goes from OFF to ON (Edge-Input), and when the
level stays ON (Level-Input).
*
“Input ON” is a status that each input terminal is shorted to the common terminal.
*
Set 50 ms or longer for the ON time and OFF time since chattering is protected.
Input is accepted only in the normal mode. In a mode like the OFF mode, input will not be
accepted to avoid wrong operations.
Control Output
*
“Output ON” is a status that each input terminal is shorted to the common terminal.
Each output terminal is turned OFF in a mode other than the normal mode to avoid wrong
operations.
Содержание Explosion Protected Weighing Indicator AD-4403-FP ST Series
Страница 22: ...2 7 Blank page...
Страница 36: ...6 3 Blank page...
Страница 38: ...6 5...
Страница 40: ...6 7...
Страница 53: ...6 20 Blank page...
Страница 73: ...8 18 Blank page...
Страница 83: ...Dimensions OP 20 Stand A 5...
Страница 85: ...A 7 Blank page...
Страница 87: ...B 2 Blank page...
Страница 105: ...Fig Apx D 02 Gravity map WMF D 2...
Страница 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...