![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315021.webp)
2-6
2-4 Installing Wall-Mounting Fittings
WARNING
The fittings have been attached temporally when shipped. Before use, affix
the wall-mounting fittings securely to the back of the unit with screws. Be
sure to apply some threadlocker to the screws after fastening the screws to
avoid loosened screws that may cause accidents.
Install the wall-mounting fittings following the instructions below.
#
Parts Name
[1] Four
screws
(M8)
[2] Wall-mounting
fittings
Placing
1: Tighten the [1] screws at a torque of
50 kgf.cm or more.
2: Apply some threadlocker to the
screws.
(Threadlocker should be type 262
manufactured by the LOCTTITE
company or an equivalent one with
90kgf.cm torque for a M8 screw)