![AND Explosion Protected Weighing Indicator AD-4403-FP ST Series Скачать руководство пользователя страница 80](http://html1.mh-extra.com/html/and/explosion-protected-weighing-indicator-ad-4403-fp-st-series/explosion-protected-weighing-indicator-ad-4403-fp-st-series_instruction-manual_2941315080.webp)
A-2
Interfaces
Input
(CONTROL IN)
•
•
•
•
6 points (6 bits, 2 Commons)
Dry contact or open collector
Signal is insulated from an internal circuit by an
optocoupler.
Terminal functions selectable by the function settings
Control I/O
Output
(CONTROL
OUT)
•
•
•
•
6 points (2 bits, 1 Common x 3)
Relay contact output
AC current output available
Rating: 24 VDC, 0.5 A (Resistance load), 220 VAC, 0.5 A
(Resistance load)
Standard serial
output(C.LOOP OUT)
Output for connection to a peripheral device manufactured by A&D (0 to 20 mA
current loop signal)
Weighing Capabilities
Zero point correcting
capability (Zero)
•
•
•
•
Clears the gross to zero in compliance with a command from the ZERO
keys, Control I/O, etc.
Capable of enabling or disabling operations when unstable.
A battery backs up zero correction value.
Adjustable range: Can be freely set within 1 to 30 % of weighing capacity.
Zero tracking capability
•
•
•
Detects gross zero point drift and performs zero correction automatically.
Tracking time: 0.0 to 5.0 (seconds) The range can be set freely.
Tracking width: 0.0 to 4.5 (D). The range can be set freely
Tare subtracting
function
•
•
•
•
Clears the net weight to zero in compliance with a command from the
TARE key, Control I/O, etc.
Capable of enabling or disabling operations when unstable or negative
weight.
The battery backs up the tare value.
Adjustable range: gross
≤
Weighing capacity
Preset tare subtracting
function
•
•
Sets/recalls presets tare for each code number.
Adjustable range: 0 < gross
≤
Weighing capacity
Stability detecting
capability
•
•
•
•
When a weighed value fluctuation amount per sampling enters a set width
within a set time, the stability mark is illuminated, judging it a stable
condition.
Capable of confirming stability through Control Output.
Stability detection time: 0.1 to 5.0 (seconds). The range can be set freely.
Stability detection width: 0 to 9 (D). The range can be set freely
Digital filter function
Integrating capability
•
•
Two digital filters connected in series
Cutoff frequency range: 0.7 to 11 (Hz)
Accumulation
capability
•
•
•
•
Accumulates the net weight automatically or manually.
The battery backs up an accumulated weight and the number of
accumulated weighing counts.
Accumulated weight range: 0 to 99999999
Accumulated weighing times range: 0 to 99999999
Содержание Explosion Protected Weighing Indicator AD-4403-FP ST Series
Страница 22: ...2 7 Blank page...
Страница 36: ...6 3 Blank page...
Страница 38: ...6 5...
Страница 40: ...6 7...
Страница 53: ...6 20 Blank page...
Страница 73: ...8 18 Blank page...
Страница 83: ...Dimensions OP 20 Stand A 5...
Страница 85: ...A 7 Blank page...
Страница 87: ...B 2 Blank page...
Страница 105: ...Fig Apx D 02 Gravity map WMF D 2...
Страница 107: ...Form F Factory setting rSF 02 5 Command mode Address None rSF 09 0 Address 1 rSF 09 1 E 2...