UG-1828
Preliminary Technical Data
Rev. PrC | Page 298 of 338
Figure 277. Main Interface of Transceiver Evaluation Software Bridge
When evaluation hardware is connected to a PC and the user wants to start using the complete evaluation system, TES will establish a
connection with the Xilinx Platform system via Ethernet connection after clicking the Connect button. When proper connection is
established, the user can configure an evaluation hardware. After selecting Connection tab, top part of that window shows the TCP IP
address (default 192.168.1.10) and Port Number (default 55557), where bottom part of the window displays information about connected
hardware and revisions of different software setup blocks. DHCP is enabled by default in Firmware version 0.14.5.5. When connecting the
evaluation platform directly to a PC the default IP address mentioned above will work. If the platform is connected via router the IP
address that is assigned will need to be identified and populated into the TES before connecting. Please contact the ADI Applications
Engineering team if the ADRV9001 Evaluation System must operate over a remote connection and a different IP address for the Xilinx
Platform is desired.
Figure 278 shows an example of correct connection between a PC and a Xilinx Platform with an ADRV9001 daughter card connected to
it. In this window user can check used hardware version as well as all software components versions used by the system in current TES
revision.