45
8 Controller Operation
ler percent output power display. Narrow the proportional band width to the
point at which the bath temperature begins to oscillate and then increase the
band width from this point to 3 or 4 times wider. Table 4 lists typical propor-
tional band settings for optimum performance with a variety of fluids at se-
lected temperatures.
Table 4
Typical Proportional Band Settings for Various Fluids
Fluid
Temperature
ProportionalBand
Stability
Water
30°C
0.31°C
±0.003°C
Water
60°C
0.31°C
±0.003°C
Eth-Gly 50%
35°C
0.31°C
±0.005°C
Eth-Gly 50%
60°C
0.31°C
±0.005°C
Eth-Gly 50%
100°C
0.4°C
±0.010°C
Oil 200, 10cs
35°C
0.6°C
±0.004°C
Oil 200, 10cs
60°C
0.6°C
±0.004°C
Oil 200, 10cs
100°C
0.6°C
±0.004°C
Oil 710
200°C
0.4°C
±0.008°C
The proportional band adjustment may be accessed within the secondary menu.
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” to access the proportional band.
S+E
Access heater power in secondary menu
Heater power in percent
S
Access proportional band
Proportional band setting
To change the proportional band press “UP” or “DOWN”.
D
Decrement display
New proportional band setting
To accept the new setting and access the cutout set-point press “SET”. Pressing
“EXIT” will exit the secondary menu ignoring any changes just made to the
proportional band value.
12 Pct
Pb=0.101C
Pb=0.060C
Содержание Kaye CTR-40
Страница 1: ...M4436 Rev B 12 19 Kaye CTR 40 Cold Temperature Reference User s Guide...
Страница 2: ......
Страница 11: ......
Страница 17: ...1 Before You Start 6 Voltage Cut In 7 5 213 247 VAC...
Страница 19: ......
Страница 27: ......
Страница 32: ......
Страница 45: ......
Страница 65: ......
Страница 73: ......
Страница 83: ...72 12 Troubleshooting 12 3 Wiring Diagram Figure 9 Wiring Diagram...