3 Specifications and Environmental Conditions
3.2
Environmental Conditions
Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
•
ambient temperature range: 5 °C to 35 °C (41 °F to 95 °F)
•
ambient relative humidity: maximum 80% for temperatures < 31°C de-
creasing linearly to 50 % at 35°C
•
pressure: 75kPa - 106kPa
•
mains voltage within ±10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitudes less than 2000 meters
•
indoor use only
3.3
Hardware Warranty and Assistance
3.3.1
Instrument Warranty
Amphenol Thermometrics, Inc. warrants its products against defects in
materials and workmanship for a period of 12 months from the date of
shipment. Amphenol Thermometrics, Inc. will, at its option, repair or replace
products which prove defective during this warranty period provided they are
returned to our facility in St. Marys Pennsylvania, European warranty returns
are sent to Pforzheim, Germany. Repairs necessitated by misuse of this product
are not covered by this warranty. No other warranties are expressed or implied,
includ- ing but not limited to the implied warranties of merchantability and
fitness for a particular purpose. Amphenol Thermometrics, Inc. is not liable for
consequential damages.
3.3.2
In-Warranty Repairs
Customers are requested to discuss their problem with a Amphenol Advanced
Sensors to insure a prompt and accurate assessment of their needs. Frequently,
a problem can be resolved via phone or FAX with minimal inconvenience or
delay.
If necessary, the Customer Service Representative will send replacement parts
or authorize the return of the instrument to the factory for repair. Instruments
Note
The Product can show some control sensitivity to moderate or severe
electromagnetic fields or conducted interference of certain frequencies. In
the presence of radiated EM disturbances, with frequencies of 250 MHz to
400 MHz and with amplitude >1 V/m to a maximum of 3 V/m, add
0.0025 °C to the stability specification. Stability is not guaranteed if
amplitude is >3 V/m. When subject to conducted disturbances of 8 MHz to
80 MHz, and amplitude >3 V, add 0.005 °C to the stability specification.
Содержание Kaye CTR-40
Страница 1: ...M4436 Rev B 12 19 Kaye CTR 40 Cold Temperature Reference User s Guide...
Страница 2: ......
Страница 11: ......
Страница 17: ...1 Before You Start 6 Voltage Cut In 7 5 213 247 VAC...
Страница 19: ......
Страница 27: ......
Страница 32: ......
Страница 45: ......
Страница 65: ......
Страница 73: ......
Страница 83: ...72 12 Troubleshooting 12 3 Wiring Diagram Figure 9 Wiring Diagram...