5 Installation
Ensure the valve handle is in the closed position before attempting to add fluid
to the tank.
Figure 1
Drain Valve Installation —IMPORTANT: Do Not Over Tighten. Follow
the installation instructions above.
5.3.2
Filling With Fluid
The CTR –40 Bath is not provided with a fluid. Depending on the desired tem-
perature range, any of the following fluids, as well as others, may be used in
the bath:
•
Water
•
Ethylene glycol/water
•
Mineral oil
•
Silicone oil
Fluids are discussed in detail in Section 7.1, Heat Transfer Fluid, on page 23.
Remove any access hole cover from the bath and check the tank for foreign
matter (dirt, remnant packing material, etc.).
Fill the bath with clean unpolluted fluid. Fill the bath carefully through the
large rectangular access hole to a level that will allow for stirring and thermal
expansion. Section 7.1.5, Thermal Expansion, explains fluid expansion. DO
NOT turn on the bath without fluid in the tank. Maximum and minimum fill
levels are indicated on the slotted baffle inside the tank. Carefully monitor the
18