15
4 Quick Start
C 25.00
Set-point 1, 25.0°C currently used
Press “SET” to select this memory and access the set-point value.
S
Access set-point value
Current value of set-point 1, 25.00°C
Press “UP” or “DOWN” to change the set-point value.
U
Increment display
New set-point value
Press SET to accept the new value and display the vernier value. The bath be-
gins heating or cooling to the new set-point.
S
Store new set-point, access vernier
Current vernier value
Press “EXIT” and the bath temperature will be displayed again.
E
Return to the temperature display
Bath temperature display
The bath heats or cools until it reaches the new set-point temperature. Turn off
the cooling to reach and control at higher temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set for
added safety. See Section 8.10, Cutout.
To obtain optimum control stability adjust the proportional band as discussed in
Section 8.9, Proportional Band.
1. 25.0
C 30.00
0.00000
24.73 C