
33
7 General Operation
The following situations benefit by changing cooling modes.
•
If the scan mode has been selected and the desired cooling scan requires
maximum cooling capacity in a range where the hot gas bypass would
normally be on.
•
The application requires maximum cooling capacity when the hot gas by-
pass would normally be on.
•
When using the bath at temperatures below 0°C in an ambient of less than
23°C, a small benefit to stability may be realized by using the hot gas by-
pass.
•
The refrigeration may be beneficially used for short times (less than one
hour) above 60°C but less than 100°C.
This list is not intended to be complete, but only suggests some of the situa-
tions when automatic modes may not be best. Most of the time, the automatic
functions are adequate and should be used.
Содержание Kaye CTR-40
Страница 1: ...M4436 Rev B 12 19 Kaye CTR 40 Cold Temperature Reference User s Guide...
Страница 2: ......
Страница 11: ......
Страница 17: ...1 Before You Start 6 Voltage Cut In 7 5 213 247 VAC...
Страница 19: ......
Страница 27: ......
Страница 32: ......
Страница 45: ......
Страница 65: ......
Страница 73: ......
Страница 83: ...72 12 Troubleshooting 12 3 Wiring Diagram Figure 9 Wiring Diagram...