25
Operation
Normal Operation
4. Configure the A SETPOINT and the B SETPOINT
To adjust the A and B setpoints which specify the upper and lower limits for the liquid
level control band, place the control mode rotary switch in the A SETPOINT position
or the B SETPOINT position, respectively. Use the RAISE/LOWER toggle switch to
adjust the respective setpoint in the same manner as described for the LENGTH
adjustment in step 2. The A and B setpoint adjustments are compatible with all three
units modes.
a. When the measured liquid level reaches or exceeds the A setpoint, the A
LEVEL LED on the front panel is energized. When the level falls below
the A setpoint, the LED is extinguished.
b. When the measured liquid level falls below the B setpoint, the B LEVEL
LED on the front panel is energized. When the level reaches or exceeds
the B setpoint, the LED is extinguished.
c. In addition to the LED functions, the controller output receptacle may be
energized and de-energized as discussed in step 5 below.
Note
The A setpoint must always be above the B setpoint. The firmware does not
allow these setpoints to be reversed. Both setpoints may be set from 0% to
100% of the LENGTH setting as long as A > B.
5. Select the operational mode of the controller output receptacle
The operation of the CONTROLLER OUTPUT receptacle of the instrument is
controlled by the fill toggle switch. Operation of the fill toggle switch is as follows:
a.
OFF:
With the power on and the fill switch in the OFF position, the
instrument serves only as a level monitor, giving a level reading on the
digital display and providing data via any analog or communication
options installed. All four setpoint LEDs (and associated J2 connector
relay contacts) operate normally, however, the controller output receptacle
on the rear panel will
always
be de-energized.
b.
ON:
With the fill switch in the ON position, the rear panel
CONTROLLER OUTPUT receptacle will become energized, thereby
initiating flow if the solenoid-operated fill valve is properly connected.
The FILL LED on the front panel will light indicating the presence of
power at the controller output receptacle.
The operator is solely
responsible for terminating the fill flow.
c.
AUTO:
With the fill switch in the AUTO position, the instrument is
capable of automatically initiating and terminating liquid fill via the
control valve, thereby maintaining the level between the selected A and B
setpoints. If the liquid level falls below the B setpoint, the rear panel
Содержание 186
Страница 3: ...2 Introduction ...
Страница 13: ...12 Installation Configuring power ...
Страница 23: ...22 Calibration Approximate Calibration ...
Страница 29: ...28 Operation Sensor contamination ...
Страница 37: ...36 RS 232 Communication Data Logger Option Error Codes ...
Страница 47: ...46 IEEE 488 Communication Option Serial Poll Status Byte ...
Страница 55: ...54 Virtual Instrument Operation Running multiple GPIB devices ...
Страница 65: ...64 Index ...