10
Installation
Installing the sensor
2. Rack mount the instrument if desired
If the instrument has a rack mount chassis, follow the following procedure:
a. Attach the rack mount adapter pieces to the instrument by first removing
the four screws on the side of the instrument that attach the cover to the
chassis. Attach the rack mount adapter pieces to the sides of the
instrument by reinstalling the screws.
b. Install the monitor in a 19" rack by securing the front panel to the rail in
each of the four corners with mounting hardware supplied by the cabinet
manufacturer.
3. Install the sensor in the cryo-vessel
Exercise care when installing the sensor since dents, crimps, bends or other physical
distortions in the thin wall capacitor will change electrical characteristics possibly
causing calibration errors and/or disruption of proper instrument operation. Before
installing the sensor, the user may want to review the
Calibration
and
Operation
sections to determine what, if any, calibration procedures may be necessary.
Note
The coaxial interconnecting cables and the oscillator are temperature
sensitive and should be mounted in such a manner as to avoid large
temperature changes such as those encountered in the path of dewar vents.
4. Connect the oscillator cable to the AMI sensor
Connect the oscillator to the sensor using a supplied 6 foot RG-59/U coaxial cable.
Ensure the oscillator is connected in the correct orientation (see page 5 for a system
diagram). The cable length between the oscillator and the sensor should not exceed 6
feet.
Caution
Moisture or contaminants in any of the BNC coaxial connectors can short out
the sensor and cause a false ‘full’ level indication or other erroneous
readings. A pack of non-conductive electrical connection lubricant (ECL) has
been included with the liquid level sensor packaging to reduce the possibility
of this occurring. If desired, apply a small amount of ECL to any of the BNC
connectors that may be exposed to moisture. Mate the doped connectors then
remove any excess ECL from the outside of the connector. Added protection
can be achieved by covering the doped connections with a short section of
heat-shrink tubing.
Note: MSDS sheets for the ECL are available upon request.
Содержание 186
Страница 3: ...2 Introduction ...
Страница 13: ...12 Installation Configuring power ...
Страница 23: ...22 Calibration Approximate Calibration ...
Страница 29: ...28 Operation Sensor contamination ...
Страница 37: ...36 RS 232 Communication Data Logger Option Error Codes ...
Страница 47: ...46 IEEE 488 Communication Option Serial Poll Status Byte ...
Страница 55: ...54 Virtual Instrument Operation Running multiple GPIB devices ...
Страница 65: ...64 Index ...