IEEE-488 Communication Option
Communication DIP Switch Settings
Communication DIP Switch Settings
The 5 DIP switches located on the rear panel of the Model 186 are used to set the
primary IEEE-488 bus address of the unit.
IEEE-488 primary bus address
The Model 186 primary bus address is controlled by switches 1 through 5 of the
communication DIP switch on the rear panel. Valid primary addresses are between 0
and 30. The Model 186 does not use secondary addressing. Note that many IEEE-488
controller cards in external computers will use address 0. The bus address for each
Model 186
should be unique
with respect to other Model 186 units or any other
devices on the bus. The switch settings for the various addresses are (on = 1 or the up
position):
DIP switch
Primary bus
address
1
2
3
4
5
off
off
off
off
off
0
off
off
off
off
on
1
off
off
off
on
off
2
off
off
off
on
on
3
off
off
on
off
off
4
off
off
on
off
on
5
off
off
on
on
off
6
off
off
on
on
on
7
off
on
off
off
off
8
off
on
off
off
on
9
off
on
off
on
off
10
off
on
off
on
on
11
off
on
on
off
off
12
off
on
on
off
on
13
off
on
on
on
off
14
off
on
on
on
on
15
on
off
off
off
off
16
on
off
off
off
on
17
on
off
off
on
off
18
1
0
1 2 3 4 5
ON
Содержание 186
Страница 3: ...2 Introduction ...
Страница 13: ...12 Installation Configuring power ...
Страница 23: ...22 Calibration Approximate Calibration ...
Страница 29: ...28 Operation Sensor contamination ...
Страница 37: ...36 RS 232 Communication Data Logger Option Error Codes ...
Страница 47: ...46 IEEE 488 Communication Option Serial Poll Status Byte ...
Страница 55: ...54 Virtual Instrument Operation Running multiple GPIB devices ...
Страница 65: ...64 Index ...