17
Calibration
Closed Dewar Calibration
If minimum and maximum liquid level indication is available via some other means
(e.g. flow calculation, visual determination, point sensors, etc.), then the presetting of
the instrument is not necessary.
Presetting the MAX/MIN calibration points
The following procedure should be performed before installation of the sensor in the
target cryo-vessel.
1. Connect the extension and oscillator cables to the J5 coaxial connector on the
rear panel of the instrument (see page 5 for a system diagram).
Do not connect
the sensor.
Energize the instrument. Press the MIN push-button through the
small hole provided on the instrument front panel. When the calibration point
has been accepted, the display will show "bbb.b" and the push-button can then
be released.
2. Connect the sensor to the oscillator cable (which is still connected to the
instrument via the extension cable). Press the MAX push-button through the
small hole provided on the instrument front panel. When the calibration point
has been accepted, the display will show "bbb.b" and the push-button can then
be released.
3. Calculate the factor
C
adj
using the following equation:
where
L
total
is the total sensor length in inches,
L
active
is the active sensor length
in inches, and
e
is the dielectric constant of the target liquid.
4. Enter
C
adj
into the instrument by placing the front panel control mode rotary
switch in the SILENCE position. By using the RAISE/LOWER toggle switch
and holding it in the up or down position, adjust the displayed value up or
down. The display will move slowly at first and then faster. Once near the
desired value, simply release the switch momentarily and then resume
changing the factor at the slower speed. Once the desired number has been
reached, release the toggle switch.
5. Once the value for
C
adj
has been entered, momentarily press the CAL push-
button labeled as "~" (the tilde character) through the small hole provided in
the instrument front panel. When the value has been accepted, the display will
show "ddd.d" and the button can then be released.
6. With the sensor connected, again press the MIN push-button through the
small hole provided on the instrument front panel. When the calibration point
has been accepted, the display will show "bbb.b" and the push-button can then
be released. The presetting procedure is complete. Proceed to the closed
dewar calibration procedure.
C
adj
120 1
2.1
L
active
5.2
L
total
------------------------
+
e
1
–
0.454
-------------
=
Содержание 186
Страница 3: ...2 Introduction ...
Страница 13: ...12 Installation Configuring power ...
Страница 23: ...22 Calibration Approximate Calibration ...
Страница 29: ...28 Operation Sensor contamination ...
Страница 37: ...36 RS 232 Communication Data Logger Option Error Codes ...
Страница 47: ...46 IEEE 488 Communication Option Serial Poll Status Byte ...
Страница 55: ...54 Virtual Instrument Operation Running multiple GPIB devices ...
Страница 65: ...64 Index ...