16
Calibration
Open Dewar Calibration
the cryogenic liquid is under pressure. For this type of approximate calibration, the
substitute
reference liquid
will be the
target liquid
at atmospheric pressure — see
page 19 for a detailed discussion of the approximate calibration method. If any
questions exist in regard to calibration issues, contact AMI for assistance in
determining the optimal calibration strategy.
Open Dewar Calibration
The instrument should be energized with the sensor connected to the instrument via
the oscillator (see the system diagram on page 5).
1. Slowly insert the sensor into the liquid until the level rests approximately one
inch below the top sensor hole and then press the MAX push-button through
the small hole provided on the instrument front panel. When the calibration
point has been accepted, the display will show "bbb.b" and the push-button
can then be released. The location of the liquid level on the sensor when the
MAX button is pressed becomes the 100% level. The 100% level should
always be lower than the upper hole to ensure the instrument will always
reach 100% in the event the overall sensor capacitance changes slightly due to
component drift, pressure variations, fluid impurities, etc.
2. Slowly withdraw the sensor out of the liquid to be measured until the level is
approximately even with the bottom hole in the sensor and then press the MIN
push-button through the small hole provided in the instrument front panel.
When the calibration point has been accepted, the display will show "bbb.b"
and the push-button can then be released. The location of the liquid level on
the sensor when the MIN button is pressed becomes the 0% level. This
completes the calibration procedure.
Note
Having a small amount of liquid on the sensor at the MIN calibration level
helps stabilize the sensor capacitance for 0% level indication.
3. Permanently install the sensor in the vessel and proceed to the
Operation
section for directions for configuring the instrument.
Closed Dewar Calibration
A calibration can be performed in a closed dewar system by monitoring the liquid
level while transferring the target liquid to an initially empty (or near empty) dewar at
a constant rate. In order to insure success with the closed dewar technique, it is
necessary to prepare the instrument by presetting the calibration MIN and MAX
points outside the estimated level range. If the instrument is not prepared in this
manner before the calibration procedure, it is possible to reach the MAX calibration
point of the instrument before the target vessel is at the desired maximum level point.
Содержание 186
Страница 3: ...2 Introduction ...
Страница 13: ...12 Installation Configuring power ...
Страница 23: ...22 Calibration Approximate Calibration ...
Страница 29: ...28 Operation Sensor contamination ...
Страница 37: ...36 RS 232 Communication Data Logger Option Error Codes ...
Страница 47: ...46 IEEE 488 Communication Option Serial Poll Status Byte ...
Страница 55: ...54 Virtual Instrument Operation Running multiple GPIB devices ...
Страница 65: ...64 Index ...