3 Operation
The Model 186 and sensor were functionally tested and calibrated at the factory. The
calibration sticker located on the bottom of the instrument shows the calibration
length, calibration liquid, and whether an approximate calibration method was utilized
at the factory. In the event that the calibration is incorrect for the application, the
instrument will need to be recalibrated by the user with a specific sensor and liquid.
Refer to the
Calibration
section for the specific procedures.
1. Energize the instrument
After completion of the
Installation
procedures, energize the instrument by placing
the power toggle switch in the POWER position. The LED display will indicate the
liquid level and the yellow ACTIVITY LED will begin blinking.
Note
The ACTIVITY LED provides visual indication that the microprocessor is
making sensor readings. If a fault should develop which prohibits the
microprocessor from operating correctly (such as a break in cabling) the
LED will not blink or blink slowly, and the display will continuously show
100%.
The instrument is normally calibrated at the factory for the specific sensor supplied
with the unit for use in a target liquid. If the need arises for recalibration, see the
Calibration
section.
2. Configure the active length setting
After calibration, the instrument
must
be configured for the active length of the sensor
in order to scale the measurement to meaningful units of length for display. For a
standard calibration, the value of the active length is the sensor length between the
bottom hole to 1 inch below the top hole of the sensor assembly. If the user performed
a calibration, then the physical distance between the locations of the MIN and MAX
calibration points on the sensor is the active length.
The instrument allows the user to display the liquid level in units of length (inches or
centimeters) in addition to a percentage. The instrument was shipped with the length
value set to the active sensor length if a sensor was purchased with the instrument.
To
view
the present length setting, place the units mode toggle switch in either the
INCH or CM position. Place the control mode rotary switch on the front panel to the
LENGTH position. Push and
release
the RAISE/LOWER toggle switch either up or
down. The display will momentarily show the current length setting.
Содержание 186
Страница 3: ...2 Introduction ...
Страница 13: ...12 Installation Configuring power ...
Страница 23: ...22 Calibration Approximate Calibration ...
Страница 29: ...28 Operation Sensor contamination ...
Страница 37: ...36 RS 232 Communication Data Logger Option Error Codes ...
Страница 47: ...46 IEEE 488 Communication Option Serial Poll Status Byte ...
Страница 55: ...54 Virtual Instrument Operation Running multiple GPIB devices ...
Страница 65: ...64 Index ...