Overview | 1-3
where R and F are constants, T is absolute temperature, and O
1
and O
2
are the oxygen partial pressures on either side of the cell.
For measuring oxygen in non-combustibles gases, the calibration of an
analyzer is obtained from the formula:
Where
A
is a constant,
T
is the cell temperature on an absolute scale (°C +
273) and O
2
Unk% is the unknown oxygen concentration of the gas to be
analyzed, and which is calculated by the analyzer.
The cell produces zero voltage when the same amount of oxygen is on
both sides, and the voltage increases as the oxygen concentration of the
sample decreases. The voltage created by the difference in the sample gas
and the reference air is carried by a cable to the microprocessor control
unit, where it is linearized to an output signal.
Because of the high operating temperature of the cell, combustible
gases that are present may burn. When this occurs, the cell will gener-
ate high millivolts and cause the display to indicate less oxygen than
is actually in the gas (net oxygen content).
Hydrocarbons
When hydrocarbons are present in the gas sample, a combustion process
occurs when this gas sample is exposed to the high temperature of the
zirconia cell. An indication that hydrocarbons may be present in the gas
sample is that the oxygen reading will be lower than expected. For exam-
ple, if a calibration gas cylinder has an oxygen value of 20 PPM and 5 PPM
of hydrogen (balance nitrogen), the oxygen analyzer will read 17.5 PPM
oxygen. The reduction of oxygen is due to the combustion process where
5 PPM of hydrogen will combine with 2.5 PPM of oxygen to form water.
Thus, the oxygen analyzer will read 17.5 PPM rather than the actual 20
PPM oxygen.
The amount of oxygen reduction is dependent on the type of hydrocar-
bons present in the sample gas.
NOTE
AT = 44.0 at 615°C
E = A*T*Log
20.9%
O
2
Unk%
NOTE
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...