Maintenance and Troubleshooting | 6-7
Thermocouple Failure
This error message occurs when the indicated sensor temperature fall
below -70° C. This error message can be the result of one of the following:
Open thermocouple
Faulty interconnecting wiring
Control unit display module failure
Verify Failure
This error message occurs if the verify gas does not read within the range
you defined. If you receive this message, calibrate the analyzer.
Zero Gas Range Error
This error message occurs if the calibration zero gas does not read within
the software’s allowable limits. Once this error occurs, calibration of the
system is aborted. This error indicates one of the following problems:
Calibration gas setup problem
Incorrect flow rates between calibrated gases
Process pressure incorrectly entered
Cell wiring reserved (Cell + and Cell - wires)
Failed cell (primary calibration should be run at this time)
Leak or plug in plumbing
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...