Controller / User Interface | 4-31
Alarm Configure
This menu option allows you to define whether alarms should trigger
based on a high alarm or a low alarm condition.
Whether Alarm 3 is set high or low applies only if it is set to track
oxygen levels (% O
2
). If Alarm 3 is set to trigger at the start of a cali-
bration or verification, this option does not apply.
If you select Hydrocarbon Upset as the Sensor Config. option from the
Setup Key menu, Alarm 4 is fixed as both a high oxygen alarm and the
alarm that triggers during a hydrocarbon upset condition. In this case,
you will not be able to select Alarm 4 as a low oxygen alarm. Alarm 3,
however, can still be selected as a high or low oxygen alarm.
Configuring the alarms
1. Select
Alarm Configure
from the Alarm menu.
2. Define Alarm 3 and Alarm 4 as high or low alarms.
3. Press the
Cancel
Key to then exit this menu option.
Relay Configure
All Series 2000 control unit relays are Form A—normally open contacts.
You can, however, use the software to set whether your alarms should
close on an alarm condition (energize on alarm) or open on an alarm
condition (de-energize on alarm). This menu option applies to all alarms,
except for the service and watchdog alarms, which are always set to de-
energize on alarm (fail-safe).
Configuring the relay
1. Select
Relay Configure
from the Alarm menu.
2. Select
De-energize on Alarm
or
Energize on Alarm
.
NOTE
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...