5-14 | Thermox TM2000 Oxygen Analyzer
The verify display then appears:
O2 xx.x Set xx.x
Press ENTER to Verify
5. Press the Enter key once the O
2
reading on the far left of the first
display line stabilizes. The reading on the far right of the first display
line shows the set point value you entered using the Configure Verify
menu option. This set point value should match the value on your
verify gas cylinder.
If your O
2
reading on the far left of the display has stabilized, yet is not
near the set point value on the right of the display, check the cal gas value
you entered and ensure that this value matches the cal gas cylinder for the
gas you applied to the cal gas inlet port of the sensor.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Verifying Cal
The system then returns to monitoring process readings.
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...