5-24 | Thermox TM2000 Oxygen Analyzer
Set Cal Timers - Cal Gas Duration
This menu option allows you to define how long the RCU should apply
the span and zero calibration gases to the sensor during an automatic
calibration or verification. For a verification, the system uses the span gas
inlet for the verify gas.
To find out how long it takes for your calibration gases to reach the sen-
sor and stabilize, select the Inject Cal Gas menu option and record the gas
stabilization period. Then use this length of time as your calibration gas
duration (we suggest that you also add a buffer to this length of time to be
certain that the gas has stabilized).
1. Select
Cal Gas Duration
from the Set Cal Timers menu option.
You will be prompted to enter the new span gas duration time:
O2 Span Time = xx:yy
Enter new time: Mn:Sc
2. Enter the new span gas duration time, then press the
Enter
key.
If performing an automatic verification, enter the verify gas time here.
Mn equals minutes and
Sc
= seconds. As you type the new span gas
duration, it overwrites the Mn:Sc text. If you wish to use the currently
displayed time, press the
Enter
key without entering a new value.
At this point you will be prompted to enter the Zero gas duration time:
O 2 Z e r o T i
me = xx:yy
Enter new time: Mn
: S c
3. Enter the new zero gas duration, then press the
Enter
key. (If perform-
ing automatic verifications, this time value is ignored since only a
single gas is used.) Mn equals minutes and Sc = seconds. As you type
the new zero gas duration time, it overwrites the Mn:Sc text. If you
wish to use the currently displayed time, press the
Enter
key without
entering any text.
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...