Controller / User Interface | 4-3
If an alarm condition exists, alarm information in the form of an Up or
Down Arrow (corresponding to a high or low alarm) will appear next to
the oxygen reading on that display line. See the Alarm Key section for de-
tails on setting alarms. If a high hydrocarbon alarm condition is reached,
an Up Arrow will appear next to the high hydrocarbon message.
Auto Ranging Feature
If the oxygen reading falls below 0.5% O
2
, the display switches from the
percent scale to the PPM scale.
Control Unit Keys
This section provides an introduction to the main Series 2000 control unit
keys. Each of these keys provides a menu of choices that can be selected.
When in a menu, enter values by pressing the number keys.
Setup
Key
Allows you to set the control unit for your specific application.
Calibrate
Key
Allows you to set calibration parameters and to start calibrations.
Alarm
Key
Allows you to define parameters that control the alarm relays.
Analog Range
Key
Allows you to define parameters that control analog current outputs.
Help
Key
Allows you to access control unit on-line help.
Cancel
Key
Cancels you out of menu options.
Enter
Key
Allows you to confirm menu choices and values.
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...