4-32 | Thermox TM2000 Oxygen Analyzer
If you select Energize on Alarm, the following message will appear:
Relays are set to
ENERGIZE on Alarm
If you select De-Energize on Alarm, the following message will appear:
Relays are set to
DE-ENERGIZE on Alarm
Select Function
This menu option allows you to select the type of reading that Alarm 3
should track:
•
Oxygen
•
When the system starts a calibration
•
When the system starts a verification
If tracking the start of a calibration or verification, the alarm will deacti-
vate once the calibration or verification has been completed, including
any recovery duration you have specified.
If you select Hydrocarbon Upset as the Sensor Config. option from the
Setup Key menu, Alarm 4 is fixed as both a high oxygen alarm and
the alarm that triggers during a hydrocarbon upset condition.
Defining the alarm function
•
Choose
Select Function
from the Alarm menu.
•
Select
Alarm3= Oxygen
,
Alarm3= In Cal.
, or
Alarm3= In Verify
, de-
pending on whether Alarm 3 should be associated with oxygen levels,
or with the start of a calibration or verification.
The control unit automatically exits you from this menu after you make
your selection.
NOTE
Содержание Thermox TM2000
Страница 12: ...xii Thermox TM2000 Oygen Analyzer This page intentionally left blank...
Страница 18: ...1 6 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 25: ...Installation and Start Up 3 3 Figure 3 1 Sensor Mounting no sample bypass...
Страница 47: ...Installation and Start Up 3 25 Figure 3 22 Current output connections example...
Страница 51: ...Installation and Start Up 3 29 Figure 3 25 Series 2000 controller RCU connections...
Страница 53: ...Installation and Start Up 3 31 Figure 3 26 Series 2000 2 wire RS 485 communications connections...
Страница 55: ...Installation and Start Up 3 33 Figure 3 27 Series 2000 4 wire RS 485 communication connections...
Страница 58: ...3 36 Thermox TM 2000 Oxygen Analyzer This page intentionally left blank...
Страница 144: ...7 8 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 162: ...B 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 166: ...C 4 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 168: ...D 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...
Страница 170: ...E 2 Thermox TM2000 Oxygen Analyzer This page intentionally left blank...