Local Operation
3-6
TM-F2OP-C1XN-01
Setting Output Voltage and Current Limit
Install the power supply and connect the load as described in
“Installation”. Ensure that the power supply is set up for local control as
described in “Shipped Configuration (Local Control Mode)” on page 3–5.
Then, set the output voltage and current limit at the front panel with the
following procedure.
1. Turn both the voltage and current controls fully counter-clockwise.
2. Turn the AC power ON.
3. Press the STANDBY switch to its IN position to disable the power
supply output. The red Shutdown (S/D) LED turns on.
4. Press and hold the V/I CHECK button to display the voltage and
current control settings on the voltmeter and ammeter displays.
5. Adjust the voltage control to the desired voltage (the compliance
voltage for applications using current mode operation).
6. Adjust the current control to the desired current limit setting.
7. Release the V/I CHECK button.
8. Press the STANDBY switch to its OUT position to apply power to the
load. The S/D LED turns off.
Note:The control circuits have been designed to allow you to set output
voltage and current up to 5% over the model-rated maximum values. The
power supply will operate within these extended ranges, but we cannot
guarantee full performance to specification.
Содержание Sorensen XFR 100-28
Страница 3: ...This page intentionally left blank ii...
Страница 5: ...This page intentionally left blank iv...
Страница 7: ...This page intentionally left blank vi...
Страница 13: ...xii...
Страница 15: ...xiv...
Страница 25: ...1 10...
Страница 67: ...3 18...
Страница 73: ...Remote Operation 4 6 TM F2OP C1XN 01 Figure 4 1 Connecting Programming Sources to J2 Connector...
Страница 75: ...4 8...
Страница 87: ...5 12...