Maintenance | 4-5
2. If the analyzer uses
measured
Pressure Compensation (optional),
change the pressure compensation to
fixed mode
so that a “
f: Sample
gas pressure too low
” alarm does not occur. To do this, view the
Cell/
Misc Parameters
dialog box and enter a value in the
Default Cell
Pres
(pressure compensation) field that is the same as, or approxi-
mately, the Measuring Cell pressure. Refer to “Cell/Miscellaneous
(Compensation) Parameters” in Chapter 4.
Close the isolation valve on the Sample Probe. Backpurge the analyzer
sample system with Zero gas for 2–3 minutes.
3. Change the
Flow Control
setting to ‘
1
’ (
shut-in/off
). Click
Apply
(next
to
Flow Control
) and then click
OK
.
Bleed down the sample system to atmospheric pressure (check gauge).
Close the valve on the Vent Line side.
4. Disconnect AC power to the analyzer and power down the computer
before performing maintenance on the analyzer.
Hazardous Locations
Always disconnect mains AC power and/or alternate power sources
to the analyzer before opening any covers or doors on the analyzer to
check or perform maintenance on any components within the enclo-
sures.
If it is necessary to open the analyzer’s covers or doors while the
circuits are live, test the area for flammable gases (and proceed only
when the area is safe). To work on the analyzer with it powered up
and its electronics enclosure(s) open, take appropriate precautions to
avoid electrical shock. Hazardous voltages are present inside.
!
WARNING
See Fuse Legend
inside analyzer’s Lower
Enclosure.
Содержание ASR900
Страница 16: ...1 4 Heated Acid Gas Sample Probe Large Diameter Figure 1 2 HAG Sample Probe assembly drawing...
Страница 25: ...Specifications 2 5...
Страница 26: ...2 6 Heated Acid Gas Sample Probe Large Diameter...
Страница 27: ...Specifications 2 7...
Страница 28: ...2 8 Heated Acid Gas Sample Probe Large Diameter...
Страница 29: ...Specifications 2 9...
Страница 52: ...3 22 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 59: ...Maintenance 4 7 Figure 4 1 Large Diameter Heated Acid Gas Sample Probe Assembly...
Страница 80: ...4 28 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 86: ...6 2 Heated Acid Gas Sample Probe Large Diameter Heated Acid Gas Sample Probe Assembly 100 2196...
Страница 88: ...6 4 Heated Acid Gas Sample Probe Large Diameter Double Block Valve Assembly 100 2393 SEE NOTES 1 2...
Страница 89: ...Drawings 6 5 Fiber Filter Assembly 100 2194...
Страница 90: ...6 6 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...
Страница 92: ...S 2 Heated Acid Gas Sample Probe Large Diameter This page intentionally left blank...